Respiratory Monitor copd-6
Model 4000
User Training Manual
Medical Devices Directive 93/42/EEC L169, Vol. 36,
EN ISO 13485, FDA QSR 21 CFR 820/803
Manufacturer: Vitalograph (Ireland) Ltd, Ennis, Ireland
UK Sales
Vitalograph Ltd.
Maids Moreton, Buckingham, MK18 1SW, England
Phone: (01280) 827110
Fax: (01280) 823302
e-mail: sales@vitalograph.co.uk
www.vitalograph.co.uk
Export Sales
Vitalograph Ltd.
Maids Moreton, Buckingham, MK18 1SW, England
Phone: +44 1280 827120
Fax: +44 1280 823302
e-mail: sales@vitalograph.co.uk
www.vitalograph.eu
Vitalograph GmbH
Rellinger Straße 64a, 20257 Hamburg, Germany
Phone: (040) 54 73 91-0
Fax: (040) 547 391 40
e-mail: info@vitalograph.de
www.vitalograph.de
Vitalograph Inc.
13310 West 99th Street, Lenexa, Kansas 66215, U.S.A.
Toll Free: 800 255 6626
Phone: (913) 730-3200
Fax: (913) 730-3232
e-mail: vitcs@vitalograph.com
www.vitalograph.com
Vitalograph (Ireland.) Ltd.
Gort Road Business Park,
Ennis, Co. Clare, V95 HFT4
Ireland
Phone: (065) 6864100
Fax: (065) 6829289
e-mail: sales@vitalograph.ie
www.vitalograph.ie
© Copyright Vitalograph 2013
Current Edition 7
Cat. No. 07381
is a registered trademark
Note: Please read all the information in this manual before using the Vitalograph copd-6 device.
*If you suspect that a defect has occurred you should check the Vitalograph copd-6 device for operation.
A |
Flowhead |
B |
On / Off button |
C |
Display |
D |
User buttons |
E |
connector |
The Vitalograph copd-6 is a device intended for measuring lung function. It can be used by a healthcare professional or assistant as a rapid pre-spirometry test to screen-out at-risk individuals that do not have COPD and indicate those that may be at risk of COPD at the presymptomatic stage of the disease. This “screening” or “case selection” for full spirometry examinations may result in earlier medical intervention and facilitate better clinical outcomes.
The Vitalograph copd-6 displays some key parameters for clinical interpretation:
If an abnormal result is indicated, arrows on the display show:
4.1 Entering Subject Data
The subject’s physical data can be entered into the Vitalograph copd-6 in order to calculate predicted data.
To enter the subject data, follow these steps:
- Turn the device on, .
- The age can now be set ( ). This is done by pressing the or button and releasing when the age is reached. The values will increase/decrease in values of 1. If the button is kept depressed, the values will scroll faster. Press the button to set the age.
- The height can now be set ( ). This is done by pressing the or button and releasing when the height is reached. The values will increase/decrease in values of 1. If the button is kept depressed, the values will scroll faster. Press the button to set the height.
Note: if height values are set below 100, the device assumes height is in inches and that Weight will be in Ibs rather than Kg.- The gender can now be set ( ). This is done by pressing the or button and releasing when the correct gender is showing. (Male ( ); Female ( )) Press the button to set the gender. After gender is selected, the Vitalograph copd-6 will go to Test Mode, where the Blow Icon is showing. On some variants, an additional data field comes first (see below).
- (This step may be omitted, depending on the variant) The Population Group (On some variants this is Weight instead) can now be set ( ). This is done by pressing the or button and releasing when the desired setting is reached. Press the button to set.
Set No. Population Group 1 NHANES III Caucasian C 2 NHANES III African-American AA 3 NHANES III Hispanic-American HA
4.2 Performing the Test
- The subject should sit down when blowing into the device (unless advised otherwise). Insert a new Bacterial Viral Filter (BVF) or disposable SafeTway mouthpiece into the device.
- Turn the device on, (if not already ON). When the device is ready for a test the blow icon shows ( ).
- Instruct the subject and demonstrate (using a mouthpiece) as follows:
- “Hold your head up, breathe in as deeply as possible, hold the device in front of your mouth”.
- “Hold your breath, place the mouthpiece into your mouth, like this, bite the mouthpiece lightly, and seal your lips firmly around it, like this”. I will demonstrate.
- “Blow out as HARD, and FAST as you can, like this, until I tell you to stop” (the unit will beep at end of test - after 6 seconds).
- “Be careful not to block the mouthpiece with your tongue or teeth. A ‘spitting’ action will give false readings”.
- “Now you do it – deep breath in – bite and seal – blast the air out…keep going – keep blowing”
- “Well done!” “Now, we need to do that three times. Rest for a while until you are ready to blow again”.
- To view the result (the best values in the session), press the button.
- The COPD Classification will show on the right hand arrow.
- Green is NORMAL, negative for COPD. No need to refer this subject for spirometry.
- Any one of the blue zones, I, II, II or IV are not normal. This subject should be referred for spirometry.
- The Obstructive Index (OI) shows on the left hand arrow.
- 0 - Green is normal.
- 1, 2 or 3 – Yellow, amber or red are not normal. Refer for spirometry, but this is unlikely to be COPD.
- Note: If the right hand (COPD) arrow is not green, nor will the left hand (OI) arrow be green.
- This is the end of the test, but if desired some test parameters may be viewed.
- Following each blow and at the end of the test session, the FEV1 value will be displayed and below that, FEV1 % Predicted results for that blow, or for the best in session if the button has been pressed. Pressing again will toggle between best and last blow.
- Pressing the button will show the FEV6 and FEV6 % Predicted results.
- Pressing the button again will show the FEV1/FEV6 and FEV1/FEV6 % Predicted results
- Press the button for a final time will show the estimated Lung Age ( )
Notes on testing:
- During testing, if an exclamation mark ! appears, this means that the last blow was not a good quality blow and the user should blow again. Reasons for poor quality are;
- Slow Start of test: Vext (extrapolated volume) is > 5% or 150mL of FEV6
- Cough detected.
- In the event that a test is < 3 seconds duration and has an abrupt end (change in volume is > 25ml in the last second of the test), the value FEV will be displayed instead of FEV6.
- If the test subject experiences side effects such as dizziness or fatigue during the test procedure, stop testing until recovered.
4.3 Setting the Obstructive Index and COPD Classification Zones
The Vitalograph copd-6 Obstructive Index and COPD Classification zones are set by the manufacturer to the GOLD (Global Initiative for Chronic Obstructive Lung Disease) standard and it is unlikely that a user will wish to change this – however if this is required please follow the following instructions:
The colour systems for each zone type are pre-set as follows:
FEV1% Pred. Obstructive Index COPD Classification FEV1/FEV6 Ratio and FEV% Pred. ≥ 80%
0 Not COPD FEV1/FEV6 > 0.7
Top Boundary Stage I FEV1/FEV6 < 0.70 and FEV1 ≥ 80% Pred. < 80% 1 Middle Boundary Stage II FEV1/FEV6 < 0.70 and FEV1 < 80% Pred. < 50% 2 Bottom Boundary Stage III FEV1/FEV6 < 0.70 and FEV1 < 50% Pred. < 30% 3
Stage IV FEV1/FEV6 < 0.70 and FEV1 < 30% Pred.
To set the boundary percentage values for both the Obstructive Index and COPD Classification zones together, follow these steps;
- Turn the device on, .
- When the icon appears, press and hold the and buttons for approximately 3 seconds.
- The top boundary can now be set. This is done by pressing the or button and releasing when the value is reached. The values will increase/decrease in values of 1%. If the button is kept depressed, the values will scroll faster.
- Press to set the top boundary value.
- The middle boundary can now be set. This is done by pressing the or button and releasing when the value is reached. The values will increase/decrease in values of 1%. If the button is kept depressed, the values will scroll faster.
- Press to set the middle boundary value.
- The bottom boundary can now be set. This is done by pressing the or button and releasing when the value is reached. The values will increase/decrease in values of 1%. If the button is kept depressed, the values will scroll faster.
- Press to set the bottom boundary value.
- Press . The device will return to the age entry screen.
4.4 Reviewing the Last Session Test Results
The Vitalograph copd-6 will always store the last test session, even after the device has powered itself down or has been switched OFF. In order to view the last test session, follow these steps:
- Turn the device on, .
- When the device is ready for age entry ( ), press the button for approximately 3 seconds. The last test session (best results) data will now show again.
- When you have finished reviewing the data, press the OFF button for 3s. OR
- Press . The device will return to the age entry screen ready for entering the next test subject’s data.
5.1 Daily Procedures
A new mouthpiece (either SafeTway or BVF) should be used for each subject. A delay of at least 5 minutes should be allowed between subjects to allow settling of previously aerosolized particles in the measuring device.
5.2 Monthly Procedures
It is recommended that the device be regularly cleaned according to the guidelines of the user’s facility. The disinfection materials and procedures applied in the users’ facility may be more appropriate than the methods outlined below. In the event of visible contamination of the flowhead element, it should be cleaned or disinfected as described in the accompanying table. The device should be replaced in the event of damage, or if visibly contaminated.
The frequency of cleaning and disinfecting is dependent on the facility’s risk assessment, usage, and test environment, but it should be at least monthly or every 100 subjects (300 blows).
5.3 Annual Procedures
It is recommended that the device be replaced annually. There is no planned preventive maintenance for this medical device.
5.4 Table of Materials Used & Cleaning/Disinfection Methods
This listing of materials used is given to provide clinical users with information to allow the assessment of other cleaning and disinfecting procedures available in the facility on this device.
Part Material Clean/ Disinfect Autoclave Possible? Recommended Disinfectants SafeTway mouthpiece or BVF Cardboard / ABS Dispose – single use No Dispose – single use
Case Exterior PC/ABS Clean No Wiping with a 70% isopropyl alcohol impregnated cloth provides a suitable form of cleaning and low-level disinfection. This may be preceded by cleaning with an anti-static foam cleaner if necessary.
Fascia PMMA/PET Clean No Removable Flowhead PC/ABS Clean No Disinfect by immersion in sodium dichloroisocyanurate solution at 1000 ppm concentration of free chlorine for 15 minutes (see following section for recommended cleaning/disinfection method for the flowhead) Maximum of 50 cleaning cycles allowed.
All external parts of the Vitalograph copd-6 require cleaning, i.e. the removal of visible particulate contamination. The parts of the device that make up the flowhead, which comes into contact with the breath of the subjects being tested, also require disinfecting. This device is not designated as a ‘sterile’ device.
5.5 Removing the Flowhead for Cleaning and Disinfecting
Flowhead Disassembled Flowhead Assembled
- Remove the flowhead from the body with a sharp pulling motion.
- Clean the flowhead by washing in a mild detergent to remove particulate contamination, taking care not to touch the moving vanes. Swill vigorously in water with mild detergent. Do not attempt to “rub” or “scrub” in the area of the vanes. Rinse with clean water.
- Disinfect by immersion in sodium dichloroisocyanurate solution at 1000 ppm concentration of free chlorine for 15 minutes. Prepare disinfectant solution as directed in the manufacturer’s guidelines. Rinse with warm water for faster drying.
- Leave it to dry completely before reassembling. Drying the flowhead may require placing it in a warm place overnight. A drying cabinet is ideal.
Wiping with a 70% Isopropyl Alcohol impregnated cloth provides a suitable form of cleaning and low-level disinfection for the case exterior, display, screen surround and keys. Repeat this at least weekly to prevent a build-up of grime from normal handling and use. Always follow the safety guidelines given by the manufacturer of cleaning and disinfectant chemicals or equipment.
Reassemble the flowhead by pushing back on until it ‘clicks’ into position. Ensure that the flowhead is pushed fully home.
Definitions of cleaning and disinfection are as defined in “Sterilization, Disinfection and Cleaning of Medical Equipment: Guidance on Decontamination from the Microbiology Committee to Department of Health Medical Devices Directorate, 1996”
Recommendations for chemical disinfectants are derived from the PHLS publication “Chemical Disinfection in Hospitals 1993"
Cat. no | Description |
---|---|
20242 | SafeTway Mouthpieces (200) |
20303 | Disposable Noseclips (200) |
20980 | Safetway Mouthpieces mini (50) |
20991 | Safetway Mouthpieces long (130) |
28350 | BVF (50) |
40167 | Pouch Spare (x10) |
Type BF equipment | |
Class II | |
The device must be taken to separate collection at the product end-of-life. Do not dispose of these products as unsorted municipal waste. | |
Attention (reference relevant section in manual) | |
Manufacturer | |
Year of Manufacture |
User Interface Symbols:
Battery status |
|
Blow Now | |
! | Bad Test (Slow start or cough) |
kglb | Unit of Weight /Measurement |
Lung Age | |
Age Icon | |
Height Icon | |
Gender Icon |
Material: |
PC/ABS |
Accuracy: |
Better than ± 3% |
Flow Impedance: |
Better than 0.15kPa/L/s at 14L/s |
Measurement Range: |
0 – 9.99 L BTPS |
Performance and Safety Standards: |
ATS/ERS Guidelines 2005 |
Electromagnetic emissions: |
CISPR 11 Group 1 (battery operated) |
Electromagnetic immunity: |
IEC 61000-4-2, IEC 61000-4-3 (battery operated) |
Sensor: |
Stator/rotor |
Power Supply: |
2 x AAA batteries |
Operating Temperature: |
17 – 37°C |
! Bad Test Criteria: |
Slow start of test (Vext>5%) or a cough detected in the first second |
FEV substituted for FEV6 | When FET < 3s and abrupt end of test |
Auto power down time |
Set to 2 minutes as standard |
Marking by the symbol compliance of the Vitalograph 4000 Respiratory Monitor copd-6 device to the Medical Devices Directive of the European Community. Such marking is indicative that the Vitalograph 4000 Respiratory Monitor copd-6 device meets or exceeds the following technical standards:
Guidance and manufacturer’s declaration – electromagnetic emissions |
||
The 4000 Respiratory Monitor copd-6 is intended for use in the electromagnetic environment specified below. The customer or the user of 4000 Respiratory Monitor copd-6 should assure that it is used in such an environment. |
||
Emissions test |
Compliance |
Electromagnetic environment – guidance |
RF emissions CISPR 11 |
Group 1 |
The 4000 Respiratory Monitor copd-6 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
RF emissions CISPR 11 |
Class B |
The 4000 Respiratory Monitor copd-6 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
Harmonic emissions |
Class A |
|
Voltage fluctuations/ |
Class A |
Guidance and manufacturer’s declaration – electromagnetic immunity |
|||||
The 4000 Respiratory Monitor copd-6 is intended for use in the electromagnetic environment specified below. The customer or the user of the 4000 Respiratory Monitor copd-6 should assure that it is used in such an environment. |
|||||
Immunity test |
IEC 60601 |
Compliance level |
Electromagnetic environment- guidance |
||
Electrostatic discharge (ESD) |
±6 kV contact ±8 kV air |
±6 kV contact ±8 kV air |
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30% |
||
Electrical fast transient/burst IEC 61000-4-4 |
±2 kV for power supply lines |
Battery operated device |
|
||
Surge |
±1kV differential mode |
Battery operated device |
|
||
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 |
<5 % 100V (>95% dip in 100V) for 0,5 cycle |
Battery operated device |
|
||
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 |
3 A/m |
Performance A |
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
Guidance and manufacturer’s declaration – electromagnetic immunity |
||||||
The 4000 Respiratory Monitor copd-6 is intended for use in the electromagnetic environment specified below. The customer or the user of the 4000 Respiratory Monitor copd-6 should assure that it is used in such an environment. |
||||||
Immunity test |
IEC 60601 |
Compliance level |
Electromagnetic environment - guidance |
|||
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3 |
10 Vrms
3 V/m |
Battery operated device
3V/m from 80MHz to 2.5 GHz |
Portable and mobile RF communications equipment should be used no closer to any part of the 4000 Respiratory Monitor asma-1 including cables, than the recommended separation distance calculated form the equation applicable to the frequency of the transmitter. Recommended separation distance d=0.35√P d=0.35√P 80MHz to 800MHz Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: |
Recommended separation distances between portable and mobile RF communication equipment and the 4000 Respiratory Monitor copd-6 |
|||
The 4000 Respiratory Monitor copd-6 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the 4000 Respiratory Monitor copd-6 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 4000 Respiratory Monitor copd-6 as recommended below, according to the maximum output power of the communications equipment. |
|||
Rated maximum W |
Separation distance according to frequency of transmitter (m) |
||
150 kHz to 80 MHz |
80 MHz to 800 MHz |
800 MHz to 2.5GHz |
|
0.01 |
0.1m |
0.1m |
0.2m |
0.1 |
0.4m |
0.4m |
0.7m |
1 |
1.2m |
1.2m |
2.3m |
10 |
3.7m |
3.7m |
7.4m |
100 |
11.7m |
11.7m |
23.3m |
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. |
Medical Devices may be affected by cellular telephones and other personal or household devices not intended for medical facilities. It is recommended that all equipment used near the Vitalograph product comply with the medical electromagnetic compatibility standard and to check before use that no interference is evident or possible. If interference is suspected or possible, switching off the offending device is the normal solution, as is required in aircraft and medical facilities.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Caution: Federal Law restricts this device to sale by, or on the order of a physician.
Product: 4000 Respiratory Monitor copd-6
Vitalograph hereby ensures and declares that the above product associated with this user manual, is designed and manufactured in accordance with the following QMS regulations and standards:
European Medical Devices Directive {MDD} 93/42/EEC
This device, classified as 2a as per Annex IX of MDD 93/42/EEC, meets the following provisions of Annex II of the Medical Devices Directive as per Article 11, section 3a, excluding point 4 of Annex II.
This device complies with the EMC Directive 89/336/EC, conformance demonstrated by following standard EN60601-1-2:2001. Equipment classification: Residential.
Canadian Medical Device Regulation {CMDR}
FDA Quality System Regulation {QSR} 21 CFR 820.
EN ISO 13485: 2012. Medical devices. Quality management systems. Requirements for regulatory purposes.
Certifying Body {for 93/42/EEC and CMDR}: British Standards Institute {BSI}
Certificate Nos. CE 00772, MD 82182, FM 83550
Signed on behalf of Vitalograph (Ireland) Ltd.
Frank Keane
General Manager, Vitalograph (Ireland) Ltd.
Subject to the conditions listed below, Vitalograph Ltd. and its associated companies, (hereinafter called the Company) guarantee to repair or at its opinion replace any component thereof, which, in the opinion of the Company is faulty or below standard as a result of inferior workmanship or materials.
The conditions of this Guarantee are: